Subject:
Handheld Radiofrequency Spectroscopy for Intraoperative Assessment of Surgical Margins during Breast-Conserving Surgery
Description:
_______________________________________________________________________________________
IMPORTANT NOTE:
The purpose of this policy is to provide general information applicable to the administration of health benefits that Horizon Blue Cross Blue Shield of New Jersey and Horizon Healthcare of New Jersey, Inc. (collectively “Horizon BCBSNJ”) insures or administers. If the member’s contract benefits differ from the medical policy, the contract prevails. Although a service, supply or procedure may be medically necessary, it may be subject to limitations and/or exclusions under a member’s benefit plan. If a service, supply or procedure is not covered and the member proceeds to obtain the service, supply or procedure, the member may be responsible for the cost. Decisions regarding treatment and treatment plans are the responsibility of the physician. This policy is not intended to direct the course of clinical care a physician provides to a member, and it does not replace a physician’s independent professional clinical judgment or duty to exercise special knowledge and skill in the treatment of Horizon BCBSNJ members. Horizon BCBSNJ is not responsible for, does not provide, and does not hold itself out as a provider of medical care. The physician remains responsible for the quality and type of health care services provided to a Horizon BCBSNJ member.
Horizon BCBSNJ medical policies do not constitute medical advice, authorization, certification, approval, explanation of benefits, offer of coverage, contract or guarantee of payment.
__________________________________________________________________________________________________________________________
As part of the treatment of localized breast cancer, breast-conserving surgery is optimally achieved by attaining tumor-free margins around the surgical resection site. Handheld radiofrequency spectroscopy for intraoperative assessment of surgical margins (eg, MarginProbe) is intended to increase the probability that the surgeon will achieve clear margins in the initial procedure, thus avoiding the need for a second surgery to excise more breast tissue.
Populations | Interventions | Comparators | Outcomes |
Individuals:
· With localized breast cancer or ductal carcinoma in situ undergoing breast-conserving surgery (lumpectomy) | Interventions of interest are:
· Handheld radiofrequency spectroscopy for intraoperative assessment of surgical margins (eg, MarginProbe) | Comparators of interest are:
· Standard surgical margin evaluation (inspection, palpation, intraoperative imaging, intraoperative histologic examination) | Relevant outcomes include:
· Change in disease status
· Morbid events |
Background
As part of the treatment of localized breast cancer, breast-conserving surgery is optimally achieved by attaining tumor-free margins around the surgical resection site. Failure to achieve clear margins will often require additional surgery to re-excise breast tissue. Currently, histologic examination of excised tissues after completion of surgery is the only method to determine definitively whether clear margins were achieved. Intraoperative methods of assessing surgical margins, such as specimen imaging, frozen section pathology, and touch print cytology, are either not highly accurate, not commonly available, or require considerable time and resources.
A device to detect positive margins should have a high sensitivity, indicating the ability to accurately detect any tumor found in the margins, ideally above 95%. While specificity is less important, excess false positive margin detection would lead to additional unnecessary tissue removal. A new device should have a specificity at least matching current standard best practices, estimated at 85%.1,
The MarginProbe is an intraperative device which uses radiofrequency spectroscopy to measure the dielectric properties of tissue into which it comes in contact. Cancer cells and normal breast tissues produce different signals. A handheld probe is applied to a small area of the lumpectomy specimen and analyzes whether the tissue is likely malignant or benign. The device gives a positive or negative reading for each touch. If any touch on a particular margin gives a positive reading, the margin is considered to be positive and more tissue should be re-excised if possible. The device can only be used on the main lumpectomy specimen; it cannot be used on shavings or in the lumpectomy cavity of the patient’s breast. Use of MarginProbe is intended to increase the probability that the surgeon will achieve clear margins in the initial surgery, thus avoiding the need for a second procedure to excise more breast tissue.
Regulatory Status
In January 2013, MarginProbe® (Dune Medical Devices, Caesarea, Israel) was approved by the U.S. Food and Drug Administration through the premarket approval process as an adjunctive diagnostic tool for identification of cancerous tissue at the margins (≤1 mm) of the main ex vivo lumpectomy specimen after primary excision. It is indicated for intraoperative use in conjunction with standard methods (eg, intraoperative imaging and palpation) for patients undergoing lumpectomy for previously diagnosed breast cancer. Food and Drug Administration product code: OEE.
Related Policies
Policy:
(NOTE: For Medicare Advantage, please refer to the Medicare Coverage Section below for coverage guidance.)
Handheld radiofrequency spectroscopy for intraoperative assessment of surgical margins during breast-conserving surgery is considered investigational.
Medicare Coverage:
There is no National Coverage Determination (NCD). In the absence of an NCD, coverage decisions are left to the discretion of Local Medicare Carriers. Novitas Solutions, Inc, the Local Medicare Carrier for jurisdiction JL, has not issued a determination for this service. Therefore, Medicare Advantage Products will follow the Horizon BCBSNJ Medical Policy.
[RATIONALE: This policy was created in 2013 and has been updated regularly with searches of the MEDLINE database. The most recent literature update was performed through January 8, 2019. The original review was informed by a 2013 TEC Assessment.1
Evidence reviews assess whether a medical test is clinically useful. A useful test provides information to make a clinical management decision that improves the net health outcome. That is, the balance of benefits and harms is better when the test is used to manage the condition than when another test or no test is used to manage the condition.
The first step in assessing a medical test is to formulate the clinical context and purpose of the test. The test must be technically reliable, clinically valid, and clinically useful for that purpose. Evidence reviews assess the evidence on whether a test is clinically valid and clinically useful. Technical reliability is outside the scope of these reviews, and credible information on technical reliability is available from other sources.
Handheld Radiofrequency for Breast Cancer margin detection
Clinical Context and Test Purpose
Breast cancer outcomes can be optimized by a thorough excision of breast cancer. A standard surgical practice of surgeons is to remove more breast tissue if pathologic examination of the initial excision shows positive margins. Handheld radiofrequency spectroscopy (for example, MarginProbe) i evaluates the resected specimen to determine if further excision is necessary during the initial lumpectomy. The use of a handheld radiofrequency spectroscopy should reduce re-excision rates, maintain low cancer recurrence rate, and minimize the volume of breast tissue excised.
The question addressed in this policy is: Does use of handheld radiofrequency spectroscopy (for example, MarginProbe) improve the net health outcomes for individuals undergoing lumpectomy for localized breast cancer or ductal carcinoma in situ (DCIS)?
The following PICO were used to select literature to inform this review
Patients
The relevant populations of interest are patients with localized breast cancer or DCIS who are undergoing lumpectomy.
Interventions
The technology being considered is handheld radiofrequency spectroscopy (for example, MarginProbe) as an adjunct to standard assessment of margins
Comparators
The following practice is currently being used: standard intraoperative assessment of margins such as inspection, palpation, intraoperative imaging, and intraoperative histologic examination. The technique used can vary by institution and surgeon. The incremental benefit of a handheld radiofrequency spectroscopy (for example, MarginProbe) may vary according to what is considered the standard intraoperative assessment.
Outcomes
The short-term outcome of interest is the re-excision rate. However, the re-excision rate can only be considered a valid outcome if long-term outcomes (eg, local recurrence rate, long-term cancer outcome) are either equivalent or in favor of handheld radiofrequency spectroscopy (for example, MarginProbe). For example, if the use of a handheld radiofrequency spectroscopy (for example, MarginProbe) results in lower re-excision rates, but local cancer recurrence rates are higher, the adequacy of the initial treatment must be questioned.
Timing
A handheld radiofrequency spectroscopy (for example, MarginProbe) is used during breast cancer surgery, with outcomes of interest including immediate re-excision rate and long-term recurrence and survival rates after cancer detection.
Setting
The setting is the outpatient surgical center.
Study Selection Criteria
Methodologically credible studies were selected using the following principles:
a. Comparative controlled prospective trials were sought, with preference for RCTs;
b. In the absence of such trials, comparative observational studies were sought, with preference for prospective studies.
c. To assess longer term outco mes and adverse effects, single-arm studies that capture longer periods of follow up and/or larger populations were sought.
d. Studies with duplicative or overlapping populations were excluded.
Technically Reliable
Assessment of technical reliability focuses on specific tests and operators and requires review of unpublished and often proprietary information. Review of specific tests, operators, and unpublished data are outside the scope of this policy and alternative sources exist. This policy focuses on the clinical validity and clinical utility.
Clinically Valid
A test must detect the presence or absence of a condition, the risk of developing a condition in the future, or treatment response (beneficial or adverse).
Pivotal Trial
The evidence evaluating the efficacy of MarginProbe comes from the pivotal trial by Allweis et al (2008) that led to device approval by the U.S. Food and Drug Administration.3,4,5,6, The reviewed trial reported the most relevant patient outcomes available for evaluating MarginProbe with the largest number of patients, including a large proportion of U.S. patients. In addition to clinical outcomes, the trial permitted assessments of diagnostic test performance of MarginProbe, which will inform judgments of its utility.
The pivotal trial (NCT00749931) compared surgical processes and short-term outcomes in patients undergoing lumpectomies for nonpalpable breast malignancies whose excised tissue was and was not assessed using MarginProbe. In both arms, surgeons could use standard of care intraoperative methods such as palpation, specimen imaging, and gross and/or microscopic pathology assessments. The pivotal trial was a multicenter (21 sites) randomized study of 596 patients assigned equally to both treatment arms. Enrolled patients met criteria described in Food and Drug Administration labeling, but all also had nonpalpable lesions that required image-guided localization. Trial design was complex and included several steps in sequence in which additional shavings of breast tissue could be taken during the operation. The principal outcome of the trial was complete surgical resection, in which positive margins were either re-excised or specifically noted if not re-excised. It was not necessary for the re-excision to result in a clear margin. This outcome is not fully clinically relevant.
For the principal outcome, surgeries using MarginProbe had a rate of successful surgical excision of 71.8% vs 22.4% for controls, with positive margin subjects as the denominator. The large magnitude of difference was statistically significant. However, this outcome was biased against the control group and included nonclinically relevant events as outcomes, such as positive margins not resected. The volume of tissue resected, on both a relative and an absolute scale, was greater in the MarginProbe group, but the trial only presents conclusions of a noninferiority analysis without specifying the noninferiority margin.
More clinically relevant outcomes included the proportion of patients with positive margins on final pathology after surgery, which was 31% for the MarginProbe group and 42% for the control group (p=0.008). Some patients with positive margins in the MarginProbe group did not have positive margins in their main specimen on final pathology. However, due to false-positive MarginProbe readings, additional shavings were taken, and cancer tissue was found at the margin. Without these additional shavings in response to MarginProbe assessment, these patients would have been considered to have clear margins. This occurrence reflects the uncertainty of final pathology in ascertaining whether all cancer tissue had been removed. The uncertainty complicated the comparison of outcomes between groups because a measure usually considered a poor outcome (eg, positive margin), in this case, was not due to inadequate surgery but to inadvertent discovery of residual cancer due to false-positive MarginProbe readings.
Re-excision rates using all patients enrolled in the trial as the denominator showed about a 5% absolute reduction in the MarginProbe group (28.5% vs 23.8%), which was not statistically significant. The decision to reoperate was based on surgeon judgment of patient and tumor characteristics and the totality of pathologic findings. The trial did not assess outcomes beyond the short-term re-excision rate; thus, it is unknown whether the lower re-excision rates resulted in at least equivalent local recurrence rates. Without knowing whether recurrence rate is at least equivalent, a lower re-excision rate could reflect inadequate initial surgery.
The trial also reported the diagnostic characteristics of MarginProbe. Of 1788 margins with final histopathology, MarginProbe readings were valid or not missing in 1750 samples. Three hundred twenty-seven margins were positive, and MarginProbe was positive in 246, for a sensitivity of 75%. Of 1423 negative margins on final pathology, MarginProbe was negative in 660, for a specificity of 46%. These performance characteristics showing moderate sensitivity and poor specificity are consistent with better-than-random capability of the device in detecting positive margins. Given the 19% (327/1750) prevalence of positive margins, the positive predictive value of a positive MarginProbe test for a margin is 24%. In another analysis (performed or requested by the Food and Drug Administration) in which the location of the positive margin was ignored and the test was considered positive if any margin tested positive, MarginProbe was 96% sensitive but only 9% specific. Although this test performance characteristic is less clinically relevant, the low specificity in this trial indicates that MarginProbe was positive for at least 1 margin in almost every patient in the trial, even though the prevalence of at least 1 positive margin was 52%.
Systematic Reviews
A systematic review by Butler-Henderson et al (2014) of techniques used for intraoperative assessment of margins in breast-conserving therapy for DCIS concluded that larger studies are needed to determine whether MarginProbe has a role to play in breast-conserving surgery.7, This conclusion was based on the pivotal trial previously reviewed and earlier studies.
A systematic review by St John et al (2017) of intraoperative techniques to assess margins (following breast conservation surgery identified 55 studies, 35 of which were included in meta-analysis.8, The primary end point was diagnostic accuracy of the various techniques, which was based on pooled sensitivity, specificity, and area under the receiver operating characteristic curve. Reviewers found only one prospective study on MarginProbe, which was found to have a diagnostic accuracy of 68.2%, based in part on sensitivity (71.4%) and specificity (67.7%). Re-excision rates were a secondary outcome: of 57 patients in the MarginProbe study, 15.8% required re-excision during the initial surgery. The MarginProbe study was not included in the meta-analysis. Other intraoperative techniques included in meta-analysis had pooled specificity ranging from 81% to 96%, depending on the modality, and pooled sensitivity ranging from 53% to 91%. The meta-analysis was limited by heterogeneity between studies in methodology and varying criteria for diagnosis and assessment of margins. A number of studies identified for the review could not be included in meta-analysis because of missing raw data.
A systematic review by Gray et al (2018) on intraoperative margin management in breast-conserving surgery identified 5 articles involving radiofrequency spectroscopy in a literature search conducted in July 2016.9, The evidence for MarginProbe showed a 70% specificity.Higher false positive rates result in higher volumes of tissue removal. When the authors considered the improved positive margin detection balanced with the limited specificity, they concluded that the routine use of MarginProbewas not recommended (grade 2B recommendation).
Nonrandomized Studies
Thill et al (2014) reported on final results of a 2011 cohort study of MarginProbe in patients with DCIS.10, 11, Forty-two (76%) of 55 patients enrolled from the general screening population at 3 centers in Germany were eligible for analysis. Patients underwent preoperative wire localization followed by breast-conserving surgery, with intraoperative assessment of the excised specimen by MarginProbe, radiograph, and paraffin-embedded pathologic review. MarginProbe also was used on additional shavings. Outcome measures were re-excision rate compared with a historical control rate of 39% and “procedure success,” defined as (1) negative margins after breast-conserving surgery and (2) early identification of an extended lesion, with conversion to mastectomy rather than re-excision. Criteria for re-excision defined a negative margin of 5 mm. The historical cohort comprised 67 patients with DCIS who underwent breast-conserving surgery by the same surgeons involved in the study during the year before enrollment began. Because information about patient selection and baseline data were not provided for either cohort, it is unknown how comparable the 2 cohorts were. Re-excision rate was 17%, a statistically significant difference from the historical control rate (p=0.018) with MarginProbe, and “procedure success” occurred in 24 (57%) of 42 patients. Sensitivity was 57% (95% confidence interval [CI], 48% to 66%) and specificity was 50% (95 CI, 42% to 58%). It is possible that the observed reduction in the re-excision rate was due to an increased incidence of mastectomies.
A retrospective, multicenter, before-after study by Sebastian et al *2015) found a reduction in re-excision procedures from 26% to 10% after introduction of Margin Probe.12, Investigators reviewed case records of 4 surgeons in 3 centers who used individual (nonstandardized), routine lumpectomy methods including criteria for re-excision (186 cases before MarginProbe; 165 cases with MarginProbe). For each surgeon, re-excision rates using MarginProbe were compared with those from a historical set, comprising a consecutive series of cases shortly before each surgeon started using MarginProbe. With the device, there were 28 cases in which the margin on the main specimen was clear, but the corresponding shaving contained cancer. Three (1.8%) of 165 patients in the “after” group underwent mastectomy; mastectomy rate in the “before” group was not reported. Performance characteristics (eg, sensitivity, specificity) of MarginProbe cannot be calculated from these data. Other study limitations included lack of baseline description of the control (“before”) group, potential confounding by secular trends over time, and lack of recurrence outcomes.
A retrospective single-center study by Blohmer et al (2016) compared the use of MarginProbe in 150 patients with a historical control group of 172 patients.13,The 2 groups had approximately similar proportions of patients with invasive breast cancer and DCIS. The historical control group underwent gross pathology examination and radiogram of the specimen as standard intraoperative procedures. The principal outcome of the study was re-excision rate. In patients for whom MarginProbe was used, the re-excision rate was 14.6%; in the historical control group, it was 29.7%. T The study did not describe the criteria for re-excision, or include long-term patient outcomes. The difference in the amount of breast tissue removed between strategies was also not reported.
A retrospective single-center study by Coble et al (2017) compared the use of MarginProbe in 137 patients with a historical control group of 199 patients.14, The 2 groups had approximately similar demographic characteristics and proportions with invasive breast cancer and DCIS. The historical control group underwent standard lumpectomy followed by additional shavings taken circumferentially from all aspects of the cavity. The principal outcome of the study was re-excision rate. For procedures using MarginProbe, the re-excision rate was 6.6%; in the historical control group, the rate was 15.1%. The total volume of tissue (main specimen plus additional shavings) removed was also less in the MarginProbe cases (78 cm3 vs 116 cm3; p= 0.002).
Kupstas et al (2017) retrospectively reviewed charts of patients from a single center who were treated with MarginProbe during lumpectomy for invasive carcinoma and DCIS; 120 patients were intraoperatively assessed using standard of care, and 120 patients were intraoperatively assessed using the MarginProbe device.15, Reviewers found an improvement in the device group for the primary outcome, re-excision rate (9.2% of patients treated with MarginProbe required re-excision surgery vs 18.2% of those treated with standard of care; p=0.039). Included in this re-excision group were those who needed a second lumpectomy 5.8% (n=7) of the device group vs 15% (n=18) of the standard care group (p=0.020). The study population differed in initial specimen volume; the device group was with significantly smaller breast volume on average (p=0.032). It also differed in the number of shavings required, as those in the device group tended to receive 1.5 more shavings than their counterparts. The final mean volume of removed tissue was comparable between the device group (53.6 mL) and the standard of care group (53.5 mL; p=0.974). Study limitations included the absence of long-term outcomes.
Clinically Useful
A test is clinically useful if the use of the results informs management decisions that improve the net health outcome of care. The net health outcome can be improved if patients receive correct therapy, or more effective therapy, or avoid unnecessary therapy, or avoid unnecessary testing.
Direct Evidence
Direct evidence of clinical utility is provided by studies that have compared health outcomes for patients managed with and without the test. Because these are intervention studies, the preferred evidence would be from randomized controlled trials.
No evidence was identified supporting the long-term utility of MarginProbe when used to assess surgical margins during lumpectomy for localized breast cancer or DCIS.
Chain of Evidence
Indirect evidence on clinical utility rests on clinical validity. If the evidence is insufficient to demonstrate test performance, no inferences can be made about clinical utility.
Current evidence does not support the clinical validity of MarginProbe, hence a chain of evidence cannot be constructed.
Section Summary: Clinically Useful
Although the nonrandomized retrospective studies showed a reduction in re-excision rate when using MarginProbe compared with historical controls, they were not rigorously controlled. Moreover, re-excision rate is an intermediate outcome that is only valid if long-term patient outcomes (eg, recurrence rate) are equivalent between MarginProbe and the alternative strategy. The single randomized controlled trial comparing short-term outcomes for patients undergoing breast surgery for nonpalpable breast malignancies managed with and without MarginProbe reported no significant difference in re-excision rates between the 2 trial arms. In addition, both the sensitivity and specificity rates for the MarginProbe were lower than those for the current standard best practices.
Summary of Evidence
For individuals who have localized breast cancer or DCIS undergoing breast-conserving surgery (lumpectomy) who receive handheld radiofrequency spectroscopy for intraoperative assessment of surgical margins (eg, MarginProbe), the evidence includes a randomized trial, several historical control studies, and a systematic review. Relevant outcomes are change in disease status and morbid events. In the randomized trial, histologic examination of surgical margins was not used in the control arm; the outcome measure (complete surgical resection) was not directly clinically relevant and was biased against the control arm; and patient follow-up was insufficient to assess local recurrence rates. The difference in re-excision rates between the 2 trial arms was not statistically significant. Diagnostic characteristics of the device showed only moderate sensitivity and poor specificity; thus, the device will miss some cancers and provide frequent false-positive results. Although several historical control studies have shown lower re-excision rates among patients in whom MarginProbe was used, the studies lacked adequate rigor to demonstrate whether the outcomes are attributable to MarginProbe. The studies did not report recurrence outcomes, which is important for assessing adequacy of resection. A randomized trial that assesses recurrence rates is required to evaluate whether the net health outcome improves with handheld radiofrequency spectroscopy compared with standard intraoperative surgical margin evaluation, including histologic techniques. The evidence is insufficient to determine the effects of the technology on health outcomes.
SUPPLEMENTAL INFORMATION
Practice Guidelines and Position Statements
National Comprehensive Cancer Network
Current National Comprehensive Cancer Network guidelines for breast cancer (v.3.2018) do not include recommendations for intraoperative assessment of surgical margins using radiofrequency spectroscopy for ductal carcinoma in situ or invasive breast cancer.16,
American Society of Breast Surgeons
The most current version of the American Society of Breast Surgeons' performance and practice guidelines for breast-conserving surgery (2015) mention that specimens should be submitted for margin assessment either intraopertively or post-surgery, depending on each institution's protocol. A recommendation for one margin assessment method over another was not made.17,
U.S. Preventive Services Task Force Recmmendations
Not applicable.
Ongoing and Unpublished Clinical Trials
Some currently unpublished trials that might impact this review are listed in Table 1.
Table1. Summary of Key Trials
| NCT No. | Trial Name | Planned Enrollment | Completion Date |
| Ongoing | | | |
| NCT02774785 | Reducing Re-excisions After Breast Conserving Surgery: A Randomized Controlled Trial Comparing the MarginProbe Device in Addition to Standard Operating Procedure Versus Standard Operating Procedure Alone in Preventing Re-excision | 460 | Dec 2018
(recruiting) |
| NCT02406599a | MarginProbe® System U.S. Post-Approval Study Protocol CP-07-001 | 440 | Jan 2019 |
NCT: national clinical trial.
a Denotes industry-sponsored or cosponsored trial.]
________________________________________________________________________________________
Horizon BCBSNJ Medical Policy Development Process:
This Horizon BCBSNJ Medical Policy (the “Medical Policy”) has been developed by Horizon BCBSNJ’s Medical Policy Committee (the “Committee”) consistent with generally accepted standards of medical practice, and reflects Horizon BCBSNJ’s view of the subject health care services, supplies or procedures, and in what circumstances they are deemed to be medically necessary or experimental/ investigational in nature. This Medical Policy also considers whether and to what degree the subject health care services, supplies or procedures are clinically appropriate, in terms of type, frequency, extent, site and duration and if they are considered effective for the illnesses, injuries or diseases discussed. Where relevant, this Medical Policy considers whether the subject health care services, supplies or procedures are being requested primarily for the convenience of the covered person or the health care provider. It may also consider whether the services, supplies or procedures are more costly than an alternative service or sequence of services, supplies or procedures that are at least as likely to produce equivalent therapeutic or diagnostic results as to the diagnosis or treatment of the relevant illness, injury or disease. In reaching its conclusion regarding what it considers to be the generally accepted standards of medical practice, the Committee reviews and considers the following: all credible scientific evidence published in peer-reviewed medical literature generally recognized by the relevant medical community, physician and health care provider specialty society recommendations, the views of physicians and health care providers practicing in relevant clinical areas (including, but not limited to, the prevailing opinion within the appropriate specialty) and any other relevant factor as determined by applicable State and Federal laws and regulations.
___________________________________________________________________________________________________________________________
Index:
Handheld Radiofrequency Spectroscopy for Intraoperative Assessment of Surgical Margins during Breast-Conserving Surgery
Radiofrequency Spectroscopy for Intraoperative Assessment of Surgical Margins during Breast-Conserving Surgery
Intraoperative Assessment of Surgical Margins during Breast-Conserving Surgery
Assessment of Surgical Margins during Breast-Conserving Surgery
MarginProbe
Margin Probe
Dune MarginProbe
References:
1. Maloney BW, McClatchy DM, Pogue BW, et al. Review of methods for intraoperative margin detection for breast conserving surgery. J Biomed Optics. 2018;23(10). PMID 30369108
2. Blue Cross and Blue Shield Association Technology Evaluation Center (TEC). Handheld radiofrequency spectroscopy for intraoperative margin assessment during breast-conserving surgery. TEC Assessments. 2013;Volume 28:Tab 4.
3. Schnabel F, Boolbol SK, Gittleman M, et al. A randomized prospective study of lumpectomy margin assessment with use of MarginProbe in patients with nonpalpable breast malignancies. Ann Surg Oncol. May 2014;21(5):1589-1595. PMID 24595800
4. Rivera RJ, Holmes DR, Tafra L. Analysis of the impact of intraoperative margin assessment with adjunctive use of marginprobe versus standard of care on tissue volume removed. Int J Surg Oncol. 2012;2012:868623. PMID 23326653
5. U.S. Food and Drug Administration (FDA). Summary of Safety and Effectiveness Data (SSED): MarginProbe System. 2012; https://www.accessdata.fda.gov/cdrh_docs/pdf11/P110014B.pdf. Accessed January 26, 2018.
6. Allweis TM, Kaufman Z, Lelcuk S, et al. A prospective, randomized, controlled, multicenter study of a real-time, intraoperative probe for positive margin detection in breast-conserving surgery. Am J Surg. Oct 2008;196(4):483- 489. PMID 18809049
7. Butler-Henderson K, Lee AH, Price RI, et al. Intraoperative assessment of margins in breast conserving therapy: a systematic review. Breast. Apr 2014;23(2):112-119. PMID 24468464
8. St John ER, Al-Khudairi R, Ashrafian H, et al. Diagnostic accuracy of intraoperative techniques for margin assessment in breast cancer surgery: a meta-analysis. Ann Surg. Feb 2017;265(2):300-310. PMID 27429028
9. Gray RJ, Pockaj BA, Garvey E, Blair S. Intraoperative margin management in breast-conserving surgery: a systematic review of the literature. Ann Surg Oncol. 2018;25:18-27. PMID 28058560
10. Thill M, Dittmer C, Baumann K, et al. MarginProbe(R)--final results of the German post-market study in breast conserving surgery of ductal carcinoma in situ. Breast. Feb 2014;23(1):94-96. PMID 24291375
11. Thill M, Roder K, Diedrich K, et al. Intraoperative assessment of surgical margins during breast conserving surgery of ductal carcinoma in situ by use of radiofrequency spectroscopy. Breast. Dec 2011;20(6):579-580. PMID 21885281
12. Sebastian M, Akbari S, Anglin B, et al. The impact of use of an intraoperative margin assessment device on re- excision rates. Springerplus. Apr 28 2015;4:198. PMID 26020017
13. Blohmer JU, Tanko J, Kueper J, et al. MarginProbe(c) reduces the rate of re-excision following breast conserving surgery for breast cancer. Arch Gynecol Obstet. Aug 2016;294(2):361-367. PMID 26796680
14. Coble J, Reid V. Achieving clear margins. Directed shaving using MarginProbe, as compared to a full cavity shave approach. Am J Surg. Apr 2017;213(4):627-630. PMID 28049561
15. Kupstas A, Ibrar W, Hayward RD, et al. A novel modality for intraoperative margin assessment and its impact on re-excision rates in breast conserving surgery. Am J Surg. Nov 21 2017. PMID 29191356
16. National Comprehensive Cancer Network (NCCN). NCCN Clinical practice guidelines in oncology: breast cancer. Version 3.2017. https://www.nccn.org/professionals/physician_gls/pdf/breast.pdf. Accessed January 24, 2018. Conserving Surgery
17. American Society of Breast Surgeons. Performance and Practice Guidelines for Breast-Conserving Surgery/Partial Mastectomy. https://www.breastsurgeons.org/statements/guidelines/PerformancePracticeGuidelines_Breast-ConservingSurgery-PartialMastectomy.pdf. Accessed January 9, 2019.
Codes:
(The list of codes is not intended to be all-inclusive and is included below for informational purposes only. Inclusion or exclusion of a procedure, diagnosis, drug or device code(s) does not constitute or imply authorization, certification, approval, offer of coverage or guarantee of payment.)
CPT*
HCPCS
* CPT only copyright 2020 American Medical Association. All rights reserved. CPT is a registered trademark of the American Medical Association.
_________________________________________________________________________________________
Medical policies can be highly technical and are designed for use by the Horizon BCBSNJ professional staff in making coverage determinations. Members referring to this policy should discuss it with their treating physician, and should refer to their specific benefit plan for the terms, conditions, limitations and exclusions of their coverage.
The Horizon BCBSNJ Medical Policy Manual is proprietary. It is to be used only as authorized by Horizon BCBSNJ and its affiliates. The contents of this Medical Policy are not to be copied, reproduced or circulated to other parties without the express written consent of Horizon BCBSNJ. The contents of this Medical Policy may be updated or changed without notice, unless otherwise required by law and/or regulation. However, benefit determinations are made in the context of medical policies existing at the time of the decision and are not subject to later revision as the result of a change in medical policy
____________________________________________________________________________________________________________________________ |